欧洲药典通论第2.5.32章水分微量测定(卡氏库仑法)修订草案已在Pharmeuropa 38.4公开征求意见,截止日期为2026年12月31日。草案为蒸发(烘箱)技术补充温度、载气流速和加热时间三项操作参数,要求背景漂移稳定且低,通常低于20 µg/min。草案还规定当各论给出温度时使用蒸发技术、未给出温度时采用直接进样,并将准确性验证章节更名为系统适用性。

正文 · 原文

A draft revision of European Pharmacopoeia (Ph. Eur.) general chapter 2.5.32. Water: micro determination has been published for public consultation in Pharmeuropa 38.4. The proposed text adds dedicated instructions on three operating parameters for the evaporation (oven) technique: temperature, carrier gas flow rate, heating time.These parameters should ensure that water is fully transferred to the titration cell without causing sample decomposition or introducing interfering substances. The briefing note states that "clarifications on the operating principle and procedure, in particular the necessity to control the background drift, as well as the applicability of reference substances for either direct introduction or evaporation techniques, have been added." Background drift should be stable and low, typically below 20 µg/min. Coulometric titration is intended for determining small quantities of water from 10 µg, while a working range of 100 µg to 10 mg is recommended for reproducibility. The draft also specifies that when an individual monograph gives a temperature, the evaporation technique is to be used. If no temperature is specified, direct sample introduction is to be used. The draft also renames the “Verification of Accuracy” section as “System Suitability” and sets recovery criteria for both direct sample introduction and evaporation. The draft text is available on the Pharmeuropa website (login required). The public consultation deadline is 31 December 2026. Comments can be submitted via the Pharmeuropa platform.

来源:ECA GMP News · gmp-compliance.org