How to Deal with Audit Trails in Preconfigured Equipment?
六位来自制药行业与检查机构的专家就预配置设备审计追踪无法配置或调整时应如何处理给出答复。若审计追踪不可配置,应先进行风险评估,确认其是否覆盖该设备的关键要求,不满足时应另选审计追踪符合要求的供应商;若无法更换,只能通过日志本和相应SOP手工记录未覆盖的操作。相关问答聚焦现行版 EU GMP Annex 11,文中提到新草案对审计追踪要求描述得更为详细。
The requirements for an audit trail can be found in all relevant regulations. The changes and deletions of data, and in future also the creation of data, should be tracked in this way. In the new draft of the EU GMP guideline Annex 11, the requirements were described in much greater detail than before. 6 Experts from the pharmaceutical industry and from inspectorates answer an extensive list of questions in this context, focusing on the currently valid Annex 11. Question 3: Most of the equipment in QC is ‘off the shelf’ and preconfigured. The audit trail cannot be adjusted or is preconfigured. How should this be handled?If the audit trail cannot be configured, a risk assessment must first be carried out to check whether it covers all the essential requirements for this device. If this is not the case, another supplier should be chosen whose audit trail meets the requirements. If this is not possible, the only solution is to manually log the uncovered actions in a logbook and a corresponding SOP. Find more Q&As on the topic 'Audit Trail' which have been answered by the expert team. The Expert Team:Klaus Feuerhelm, Local GMP Inspectorate / Regierungspräsidium TübingenEberhard Kwiatkowski, PharmAdvantageIT, NeuschooDr Ullrich Opitz, Merck, DarmstadtYves Samson, Kereon, BaselDr Wolfgang Schumacher, ehem. F. Hoffmann-La Roche, BaselDr Arno Terhechte, Local GMP Inspectorate / Bezirksregierung Münster